5.3.2

Electrode assembly installation

!

Caution

Risk of minor or moderate injury by electric shock.
• Switch off the instrument before connecting or disconnecting the high-voltage cable
• Do not touch the high-voltage cable with wet hands
• Use only damage free cabling

NOTE
Before installing the electrode, clean all parts
a to e thoroughly.

a Position the flat seal at the particle collector

basement.

b Position the metal collecting tube electrode

on the collector basement and assure elec-
trical contact. The broader uncoated stripe
at the lower outer side is to be installed
towards the base.

c Install the insulation glass cylinder on the

collector basement.

d Place the assembly on the lifting device.