8 Troubleshooting

88

B-90 Operation Manual, Version B

Table 8-1: Error messages and their remedy (cont.)
011

Inlet temperature exceeded

Heating control or heating relais
defective

Switch off the device and try
again. If the error still occurs
replace the heater module or
contact the BUCHI customer
service.

014

Voltage HV-Module

HV-module defect or internal
wiring interruption or quality fall-
off voltage because of a short
circuit

Switch off the device and try
again. If the error still occurs
contact the BUCHI customer
service.

015

Current in HV-Module too high

HV-module defect or internal
wiring interruption

Switch off the device and try
again. If the error still occurs
contact the BUCHI customer
service.

016

Current in HV-Module too low

HV-connector not plugged in or
defective wiring error

Switch off the device and try
again. Check the HV-module,
connection. If the error still
occurs , contact the BUCHI
customer service.

020

Relative pressure sensor 1
not connected

Sensor not connected or defec-
tive internal wiring

Contact BUCHI customer service

021

Relative pressure sensor 2
not connected

Sensor not connected or defec-
tive internal wiring

Contact BUCHI customer service

023

Difference pressure sensor
not connected

Sensor not connected or defec-
tive internal wiring

Contact BUCHI customer service
BUCHI

024

Inside pressure low

System not leakproof

Test for leaks

025

Filter pressure drop low

Filter not installed or defective
internal wiring

Check if outlet filter is blocked.
Replace the filter.

026

Flow rate low

No flow rate, system not leak-
proof

Generate gas flow, test for leaks

027

Inside pressure high

Gas pressure connection too
high

Regulate gas pressure

028

Flow rate high

Sensor is at stop position

Reduce gas flow rate

029

Relative pressure sensor 1
defective

Sensor or internal wiring defec-
tive

Contact BUCHI customer service

030

Relative pressure sensor 2
defective

Sensor or internal wiring defec-
tive

Contact BUCHI customer service

032

Difference pressure sensor
defective

Sensor or internal wiring defec-
tive

Close gas flow rate and restart
the system, contact BUCHI
customer service

035

No glass isolation cylinder

Position switch

Install glass insolation cylinder

040

Valve 1 CO

2

not connected

Internal wiring defective or
missing

Contact BUCHI customer service

041

Valve 2 O

2

not connected

Internal wiring defective or
missing

Contact BUCHI customer service

042

O

2

sensor not connected

Defective sensor, sensor cable or
internal wiring

Contact BUCHI customer service