beautypg.com

8troubleshooting, 1 error messages and their remedy – BUCHI Nano Spray Dryer B-90 User Manual

Page 87

background image

8 Troubleshooting

87

B-90 Operation Manual, Version B

8

Troubleshooting

This chapter helps to resume operation after a problem has occurred with the instrument which does
not require special technical training. It lists possible occurrences, their probable cause and suggests
how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Remedy”.
Malfunctions or errors which are not listed in the table are to be corrected by a BUCHI trained tech-
nical engineer who has access to the official service manuals. In this case, please refer to your local
BUCHI customer service agent.

8.1

Error messages and their remedy

Table 8-1: Error messages and their remedy

Error number Error reading

Possible cause

Remedy

001

Temperature sensor
spray head not connected

Defective sensor, sensor cable or
internal wiring

Check if the spray head plug is
plugged in correctly. Switch off
the device and try again. If the
error still occurs replace the
spray head or contact the BUCHI
customer service.

003

Resonance not found
check the spray head

No spray nut, spray nut not tight
enough, spray head defective

Switch off the device and try
again, if the error still occurs
clean the spray cap, replace the
spray cap, replace the spray
head or contact the BUCHI
customer service.

005

Spray head temperature
exceeded

Piezo defective, temperature
problem

Switch off the device and try
again. If the error still occurs
replace the spray head or
contact the BUCHI customer
service.

007

Temperature sensor
outlet not connected

Defective sensor, sensor cable or
internal wiring

Switch off the device and
try again. Check if the outlet
temperature sensor is plugged
in correctly. If the problem
still occurs replace the outlet
temperature sensor or contact
the BUCHI customer service.

008

Temperature sensor
inlet not connected

Defective sensor, sensor cable or
internal wiring

Switch off the device and
try again. Check if the inlet
temperature sensor is plugged
in correctly. If the problem
still occurs replace the inlet
temperature sensor or contact
the BUCHI customer service.