beautypg.com

BUCHI Nano Spray Dryer B-90 User Manual

Page 42

background image

5 Putting into operation

42

B-90 Operation Manual, Version B

All nuts must be tightened
crosswise only!

4

c Mount the inlet temperature sensor at the heater

module. Keep the temperature sensor end about
70 mm out of the screw connection. Use a flat
spanner to tighten the screw fitting.

d Place the heater module on the metal screw bars.

NOTE
• Position four hex nuts from top onto the metal

bar ends with a narrow gap towards the heater
module.

• Position the hex nuts from below the heater with

approximately the same narrow gap towards the
heater module.

5

—

Make sure the spray cylinder is in-line with the heater
module and that the sealing is properly placed
between.
e–f Tighten the hex counter nuts above and below

the heater crosswise. Do not overtighten the
nuts.

g Remove the mounting tool from the lifting device

after installation.

Final steps:
• Connect the electrical cable of the heater to the

back side of the instrument (see section 4.10,
socket m).

• Connect the inlet temperature sensor to its socket

at the back side of the instrument (see section 4.10,
socket l).

• Connect the drying gas tubing to the inlet of the

heater module. Use a hose clamp for fixation.

• Connect the drying gas tubing between the heater

inlet and the connector nipple inside the instrument
(open the upper back panel of the instrument to get
access).