Viscometer Will Not Return to Zero

✓ Rheometer is not level

• Adjust the laboratory stand

✓ Pivot point or jewel bearing faulty

• Perform calibration check
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Inaccurate Readings

✓ Verify spindle, speed and model selection.

✓ If % readings are under-range (less than 10%), the units display (%, cP, D/cm

2

, 1/sec) will

flash; change spindle and/or speed.

✓ “EEEE” on the digital display means the unit is over-range (greater than 100%); reduce speed

and/or change spindle.

✓ Verify test parameters: temperature, container, volume, method. Refer to:

• “More Solutions to Sticky Problems”, Section III
• DV-III+ Digital Rheometer Operating Instructions, Appendix C, “Variables in Viscosity

Measurements.”

✓ Perform a calibration check; follow the instructions in Appendix E.

✓ Verify tolerances are calculated correctly.

✓ Verify the calibration check procedures were followed exactly.

If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix I for
details on “How to Return Your Rheometer.”

Display Reading Will Not Stabilize

✓ Special characteristic of sample fluid. There is no problem with the rheometer.

• Refer to Appendix C

✓ Check for erratic spindle rotation

• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓ Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓ Temperature fluctuation in sample fluid

• Use temperature bath control.