Program 4:

Spring relax

Command

Command Description

Parameter

Comments

WKY

Wait for a key press

Wind to 100%

tell operator to wind spindle
until 100% torque is reached

WPT

Wait for % torque

100.0

wait until 100% torque is
reached

WKY

Wait for a key press

Release spindle

tell operator to release the
spindle

SPI

Set print interval

00:01

begin printing data at 1 second
intervals

WPT

Wait for % torque

0.0

wait for spindle to completely
unwind to 0% torque

Program 5:

Variable speed

Command

Command Description

Parameter

Comments

SSN

Set viscometer speed

5.0

run at 50 RPM

WTI

Wait for time interval

00:10

wait for 10 seconds

PDN

Print data point now

print one data point

SSN

Set viscometer speed

10

run at 10 RPM

WTI

Wait for time interval

00:10

wait for 10 seconds

PDN

Print data point now

print one data point

SSN

Set viscometer speed

20

run at 20 RPM

WTI

Wait for time interval

00:10

wait for 10 seconds

PDN

Print data point now

print one data point

SSN

Set viscometer speed

5.0

run at 50 RPM

WTI

Wait for time interval

00:10

wait for 10 seconds

PDN

Print data point now

print one data point