beautypg.com

Brookfield DV-III Rheometer User Manual

Page 27

background image

Brookfield Engineering Laboratories, Inc.

Page 27

Manual No. M/98-211-A0701

This screen advises that there are 7 speed sets in memory; speed set #3 is in use and that the DV-
III+
is awaiting input for the speed set to delete.

Note:

If no speed set is in use the word “NONE” will appear next to the IN USE: prompt.

At this point, the user has two options:

1. Pressing the MOTOR ON/OFF/ESCAPE key will exit from this screen and no speed sets will

be cleared. Or

2. Pressing any of the keys “0”, “1”, “2”, “3, 5, 8, or “9” will delete that speed set.

Thus, to discard speed set #5, the user would press the “5 key and be presented with:

Figure 31

In which the DV-III+ is requesting that the user specifically press the “1 key in order to delete the
desired speed set.

Figure 32

Pressing the “3” key will cause the DV-III+ to take no action will return the user to the CLEAR
SPEED
SET opening screen, Figure 30. Any attempt to delete an in-use speed (“3” for instance)
will cause the DV-III+ to issue a “beep” “beep” with no action being taken. Thus no active (i.e.
selected for use) program can be deleted from this screen.

USING PRE-PROGRAMMED SPEEDS

Pressing the three (3) key from Figure 21 takes the user to the speed set selection screen of Figure
22
where the user selects a new speed set. That done the user is sent to the default screen with:
SPEED SET X SELECTED” displayed on line three (3) of the screen.

The user initiates the use of programmed speeds by pressing the

PROG RUN

key. If the user presses

the

PROG RUN

key with no speed set selected, the following error box will be displayed: