V.4 Set Temperature

The DV-III+ can issue temperature control commands when an external temperature control device
has been connected (see Section V.1). Selecting item #4 Set Temp in the Options menu will display
the current setpoint. Enter the new setpoint by using the number keys and accept with the

ENTER

key. The temperature controller will begin using the new setpoint immediately upon the press of
the

ENTER

key.

Note:

The sign of temperature is changed by using the arrow keys when the cursor is
under the

±

character. This is possible only with temperature bath controllers.

V.5 Data

The Review Data menu allows data review of the most recently completed speed set program. Data
may be reviewed on the DV-III+ screen, on a serial printer, or both. The procedure for reviewing
data after a test is as follows:

1) Run a DV-III or B.E.V.I.S. program.

2) Press the

OPTION/TAB

key.

3) Press the

NUMBER 5

key to display the Review Data screen.

REVIEW DATA
1 = SCREEN ONLY
2 = SCREEN & PRINT
3 = PRINTER ONLY

Figure 62

4) Select one of the review modes: press “

NUMBER 1

” for DV-III+ screen only, “

NUMBER 2

” for

DV-III+ screen and printer, or “

NUMBER 3

” for printer only. Note that you should select options

2 and 3 only if the printer is connected and “on-line.”

5) If

NUMBER 1

or

NUMBER 2

were pressed, data from the first step is either displayed on the screen

or displayed and printed. Press the

ENTER

key to scroll through and display/print the remaining

step data.

cP=10800

STEP01

%=50.2 TIME=00:00
SS=0.0 TEMP=70.2
SR=0.0 RPM=100.0

Figure 63

6) If the

NUMBER 3

key (printer only) is selected, the DV-III+ prompts to

READY PRINTER PRESS

ANY KEY” or “PRINTER IS NOT READY TURN ON/PRESS ONLINE

. A sample of printed output is

shown in Figure 64.