Brookfield DV-III Rheometer User Manual
Page 14
Brookfield Engineering Laboratories, Inc.
Page 14
Manual No. M/98-211-A0701
T i m e
XXX.X
M
S
T
M o d e l
Spindle
T o r q u e
V i s c o s i t y
Shear Stress
cP or mPas
D/CM2 or
N/M2 =
D / C M 2 / 1 0
XXXXX
XX
X.XX
XXX.X
XXXXX
XX.X
0 <= cP <= 9.99
10 <= cP <= 999.9
1000 <= cP <= 99999
00 <= S <= 99
-10.1 <= T <= 99.9
See Model Table D2 - App D
12345
RV
31
82.4
3.16
123.8
XXXeX
100000 <= cP <= 51200000000 123e3 to 8
4.56
12345
0 <= D/CM2 <= 9.99
X.XX
XXX.X
XXXXX
10 <= D/CM2 <= 999.9
1000 <= D/CM2 <= 99999
234.5
Shear Rate
1 / S E C
T e m p e r a t u r e
T
-99.9 <= T <= 300.0
-10.3
X.XXX
0 <= 1/SEC <= 9.999
XX.X
XXXXX
10 <= 1/SEC <= 99.9
100 <= 1/SEC <= 99999
1.234
20.7
200
0.01 <= RPM <= 0.99
R P M
R P M
I t e m
P r i n t
F o r m a t
Example
Range
0.09
X.XX
XX.X
0.1 <= RPM <= 99.9
2.4
XXX.X
100 <= RPM <= 250
150.0
XXXeX
100000 <= D/CM2 <= 999999
123e3
Z
XX:XX
00:00 <= Z
<= 99.59
05:32
T i m e
Table 1
OUT OF RANGE INDICATORS
The DV-III+ is capable of measuring instrument torque within the range of 0 to 100%. Based on
this measurment, viscosity and shear stress are calculated. Brookfield recommends that data be
collected only in the range of 10 to 100%. Any data collected outside of this range is considered
invalid.
The DV-III+ provides the following display indicators when the measurement point is outside of
the 10-100% acceptable range.
TORQUE GREATER THAN 100%
When Rheometer torque exceeds 100%, the parameter display field will show “EEEE” for torque,
viscosity and shear stress.
Figure 8
TORQUE LESS THAN 10%
When Rheometer torque drops below ten (10) percent, the Rheometer will continue to display
measurement (%, cP, D/cm
2
) values with units flashing: