Figure 65

Note:

If continuous printing was used during the test, it will be suspended when viewing the
test data or entering time to torque parameters.

TIMED STOP

The DV-III+ will operate at a single RPM for a specified period of time. Time and speed are entered

using the number keys and

OPTIONS/TAB

key. The test will begin immediately upon the press of

the

ENTER

key.

The test will end when the specified time interval has elapsed. The data will be shown as displayed

in Figure 66. The

SELECT DISPLAY

key can be used to view all measurement parameters. The

PRINT

key can be used to send a single dataline to the printer.

Figure 66

Note:

Continuous print mode may be used in conjunction with Timed Stop to print data
throughout the time period. Printing will occur only during the test.

TIMED AVERAGE

The DV-III+ will collect a specified number of data points over a period of time and present the

average reading (arithmetic average). Time, speed and the number of data points are entered using
the number keys and

OPTIONS/TAB

key. The test will begin immediately upon the press of the

ENTER

key. Non-averaged data will be displayed during the test.

The test will end when the specified time interval has elapsed. The averaged data point will be

shown as displayed in Figure 67. The

SELECT DISP

key can be used to view the average value of

all measurement parameters. The

PRINT

key can be used to send a single averaged dataline to the

printer.