If continuous printing was used during the test, it will be suspended when viewing the
test data or entering time to torque parameters.
TIMED STOP
The DV-III+ will operate at a single RPM for a specified period of time. Time and speed are entered
using the number keys and
OPTIONS/TAB
key. The test will begin immediately upon the press of
the
ENTER
key.
The test will end when the specified time interval has elapsed. The data will be shown as displayed
in Figure 66. The
SELECT DISPLAY
key can be used to view all measurement parameters. The
key can be used to send a single dataline to the printer.
Figure 66
Note:
Continuous print mode may be used in conjunction with Timed Stop to print data
throughout the time period. Printing will occur only during the test.
TIMED AVERAGE
The DV-III+ will collect a specified number of data points over a period of time and present the
average reading (arithmetic average). Time, speed and the number of data points are entered using
the number keys and
OPTIONS/TAB
key. The test will begin immediately upon the press of the
ENTER
key. Non-averaged data will be displayed during the test.
The test will end when the specified time interval has elapsed. The averaged data point will be
shown as displayed in Figure 67. The
SELECT DISP
key can be used to view the average value of
all measurement parameters. The
key can be used to send a single averaged dataline to the
printer.