Brookfield DV-III Rheometer User Manual
Page 64
Brookfield Engineering Laboratories, Inc.
Page 64
Manual No. M/98-211-A0701
Table D1 (continued)
Table D2 lists the model codes and spring torque constants for each Rheometer model.
Table D2
The full scale viscosity range for any DV-III+ model and spindle may be calculated using the equation:
where: Full Scale Viscosity Range [cP] = TK * SMC *
10,000
RPM
TK
=
DV-III+ Torque Constant from Table D2
SMC
=
Spindle Multiplier Constant listed in Table D1
SC4-27
27
25
0.34
SC4-28
28
50
0.28
SC4-29
29
100
0.25
SC4-31
31
32
0.34
SC4-34
34
64
0.28
SC4-37
37
25
0.36
CP40
40
0.327
7.5
CP41
41
1.228
2
CP42
42
0.64
3.8
CP51
51
5.12
3.84
CP52
52
9.83
2
Spiral Adapter
70
105
0.677
Thermosel DIN spindle
81
3.7
1.29
SSA DIN spindle
for 13R or 13RP chamber
82
3.75
1.29
SSA DIN spindle
for 7R or 7RP chamber
83
12.09
1.29
ULA DIN spindle
85
1.22
1.29
ULA DIN spindle
86
3.65
1.29
ULA DIN spindle
87
12.13
1.29
SPINDLE
ENTRY CODE
SMC
SRC
LVDV-III+
0.09373
LV
2.5xLVDV-III+
0.2343
4L
5xLVDV-III+
0.4686
5L
1/4 RVDV-III+
0.25
1R
1/2 RVDV-III+
0.5
2R
RVDV-III+
1
RV
HADV-III+
2
HA
2xHADV-III+
4
3A
2.5xHADV-III+
5
4A
HBDV-III+
8
HB
2xHBDV-III+
16
3B
2.5xHBDV-III+
20
4B
5xHBDV-III+
40
5B
MODEL
TK
MODEL CODE
ON DV-III SCREEN