beautypg.com

Brookfield DV-III Rheometer User Manual

Page 64

background image

Brookfield Engineering Laboratories, Inc.

Page 64

Manual No. M/98-211-A0701

Table D1 (continued)

Table D2 lists the model codes and spring torque constants for each Rheometer model.

Table D2

The full scale viscosity range for any DV-III+ model and spindle may be calculated using the equation:

where: Full Scale Viscosity Range [cP] = TK * SMC *

10,000

RPM

TK

=

DV-III+ Torque Constant from Table D2

SMC

=

Spindle Multiplier Constant listed in Table D1

SC4-27

27

25

0.34

SC4-28

28

50

0.28

SC4-29

29

100

0.25

SC4-31

31

32

0.34

SC4-34

34

64

0.28

SC4-37

37

25

0.36

CP40

40

0.327

7.5

CP41

41

1.228

2

CP42

42

0.64

3.8

CP51

51

5.12

3.84

CP52

52

9.83

2

Spiral Adapter

70

105

0.677

Thermosel DIN spindle

81

3.7

1.29

SSA DIN spindle

for 13R or 13RP chamber

82

3.75

1.29

SSA DIN spindle

for 7R or 7RP chamber

83

12.09

1.29

ULA DIN spindle

85

1.22

1.29

ULA DIN spindle

86

3.65

1.29

ULA DIN spindle

87

12.13

1.29

SPINDLE

ENTRY CODE

SMC

SRC

LVDV-III+

0.09373

LV

2.5xLVDV-III+

0.2343

4L

5xLVDV-III+

0.4686

5L

1/4 RVDV-III+

0.25

1R

1/2 RVDV-III+

0.5

2R

RVDV-III+

1

RV

HADV-III+

2

HA

2xHADV-III+

4

3A

2.5xHADV-III+

5

4A

HBDV-III+

8

HB

2xHBDV-III+

16

3B

2.5xHBDV-III+

20

4B

5xHBDV-III+

40

5B

MODEL

TK

MODEL CODE

ON DV-III SCREEN