Din adapter, Spiral adapter, Cone/plate rheometer – Brookfield DV-III Rheometer User Manual
Page 59: Helipath with t-bar spindles
Brookfield Engineering Laboratories, Inc.
Page 59
Manual No. M/98-211-A0701
Din Adapter
Shear
DAA
Rate
Spindle
(sec-1)
LVDV-III+
RVDV-III+
HADV-III+
HBDV-III+
Viscosity (cP)
85
0 - 258
0.6 -
5,000
6.1 -
5,000
12.2 -
5,000
48.8 - 5,000
86
0 - 258
1.8 - 10,000
18.2 - 10,000
36.5 - 10,000
146
- 10,000
87
0 - 258
5.7 - 50,000
61.0 - 50,000
121
- 50,000
485
- 50,000
Spiral Adapter
Spiral
Shear Rate
Spindle
(sec-1)
LVDV-III+
RVDV-III+
HADV-III+
HBDV-III+
SA-70
.677 - 67.7
98.4 - 98,416 1,050 - 1,050,000
2,100 - 2,100,000
8,400 - 8,400,000
(1 - 100 RPM
)
Viscosity (cP)
Cone/Plate Rheometer
Shear
Cone
Rate
Spindle
(sec-1)
LVDV-III+C/P
RVDV-III+C/P
HADV-II+IC/P
HBDV-II+IC/P
Viscosity (cP)
CPE-40
0 - 1500 0.15 - 3,065
1.7 -
32,700
3.3 -
65,400
13.1 -
261,000
CPE-41
0 -
400 0.58 - 11,510
6.2 - 122,800
12.3 -
245,600
49.1 -
982,400
CPE-42
0 -
760 0.3
- 6,000
3.2 -
64,000
6.4 -
128,000
25.6 -
512,000
CPE-51
0 -
768 2.4
- 47,990 25.6 - 512,000
51.2 -1,024,000
205.0 -4,096,000
CPE-52
0 -
400 4.6
- 92,130 49.2 - 983,000
98.3 -1,966,000
393.0 -7,864,000
Helipath with T-Bar Spindles
T-Bar
Spindle
LVDV-III+
RVDV-III+
HADV-III+
HBDV-III+
T-A
156 -
187,460
2,000 -
200,000
4,000 -
4,000,000
16,000 - 16,000,000
T-B
312 -
374,920
4,000 -
400,000
8,000 -
8,000,000
32,000 - 32,000,000
T-C
780 -
937,300
10,000 - 10,000,000
20,000 - 20,000,000
80,000 - 80,000,000
T-D
1,560 - 1,874,600
20,000 - 20,000,000
40,000 - 40,000,000
160,000 - 160,000,000
T-E
3,900 - 4,686,500
50,000 - 50,000,000 100,000 -100,000,000
400,000 - 400,000,000
T-F
7,800 - 9,373,000 100,000 -100,000,000 200,000 -200,000,000
800,000 - 800,000,000
Viscosity (cP)
In taking viscosity measurements with the DV-III+ Rheometer there are two considerations which
pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range from
10% to 100% for any spindle/speed combination.
2) Viscosity measurements should be taken under laminar flow conditions, not under turbulent
flow conditions.
The first consideration has to do with the precision of the instrument. All DV-III+ Rheometers have a