Appendix j - dve-50a probe clip – Brookfield DV2T Viscometer User Manual
Page 86
Brookfield Engineering Labs., Inc.
Page 86
Manual No. M13-167-B0614
Appendix J - DVE-50A Probe Clip
Probe Clip DVE-50A is supplied with all model DV2T Viscometers, DV-III Rheometers,
and Digital Temperature Indicators. It is used to attach the RTD temperature probe to
the LV Guard Leg (Part No. B-20Y) or 600 mL low form Griffin beaker. Figure J-1 is a
view of the Probe Clip, showing the hole into which the RTD probe is inserted, and the
slot which fits onto the LV guard leg. When inserting the RTD probe into the Probe Clip,
the upper part of the Clip is compressed by squeezing the points shown in Figure J-1.
Note
:
All Viscometer/Rheometer models — except LV — use the
Probe Clip as shown in Figure J-3.
RTD Probe
Hole
Squeeze Here when
installing RTD
Temperature probe
Figure J-1
Figure J-2 shows the Probe Clip (with RTD temperature probe installed) mounted on
the LV guard leg.
Figure J-3 shows the Probe Clip mounted in a 600 mL low form Griffin beaker. This
mounting may be used with LV, RV, HA and HB series instruments.
Figure J-2
Figure J-3
Brookfield 600 mL Low Form
Griffin Beaker
DVE–50
Probe Clip
RTD Temperature Probe
LV Guard Leg
Assembly B-20KY
Probe Clip
DVE-50
R TD Temperatur e
Probe
Caution: Temperature probe must not contact the spindle during
measurement.
Note
:
The RTD probe must be parallel to the beaker wall so as not to
interfere with the viscosity measurement.