Brookfield DV2T Viscometer User Manual
Page 42
Brookfield Engineering Labs., Inc.
Page 42
Manual No. M13-167-B0614
The Oscillation Check is not a guarantee of proper calibration. It is only
an indication of the performance of the lower bearing of the viscometer.
Calibration can only be verified through the use of a calibration standard
such as Brookfield Silicone and Mineral Oil standards.
Technical Support Info:
The information contained in the Technical Support Info menu
is designed to support Brookfield for problem resolution. When
contacting Brookfield for support, you may be asked to provide
information contained in one of the menu choices within Technical
Support Info. There are no settings within this section and the
performance of the DV2T Viscometer can not be affected by
accessing this section.
About:
The About screen is shown during the start up sequence of the DV2T Viscometer.
The user can see the About screen at any time by selecting About in the Device
Setup menu (see Figure IV-4). Information in the About screen includes: Torque
Range (LV, RV, HA, HB), Model Number, Firmware Version, Viscometer Serial
Number, TK version (torque table), Spindle version (spindle table).
Figure IV-4
TIP: The information on the About screen will be needed when contacting
Brookfield for support.
To return from the About screen, press the Left Arrow navigation button in the Title
Bar.
IV.2 User Settings
The User Settings menu includes settings related to specific users. The DV2T may be
set up for a single user setting (used by one or more operators) or for multiple users who
access through a Log In and Password. Three User settings are available including: