Brookfield DV2T Viscometer User Manual
Page 28
Brookfield Engineering Labs., Inc.
Page 28
Manual No. M13-167-B0614
Collect Single Point at Step End. During the test, the total time required to
reach 200 cP is 10 minutes 40 seconds. Total points collected will be 11
with the last data point taken at 10 minutes 40 seconds, an average of the
last 20 seconds of the step.
No Data
: End Condition is met and no data is collected.
III.6 End Condition
The completion of a test is defined by the End Condition. Each time that you enter Configure
Viscosity Test, the End Condition will be set to the last value used. The End Condition
parameter or values can be changed by pressing the End Condition button.
Within the End Condition screen, the currently selected End Condition parameter and
values are displayed. The End Condition can be changed by pressing the blue down
arrow within the parameter field. Six End Conditions are available: Time, # of Points, # of
Revolutions, Torque, Viscosity, Temperature (see Figure III-5).
Figure III-5
Time
: The test will complete when the specified amount of time has elapsed. Time is
entered in Hours, Minutes, and Seconds.
Hours: 0-99
Minutes: 0-59
Seconds: 0-59
A time of zero hours, minutes and seconds may be selected. With this End
Condition, the DV2T will operate at the selected speed until the operator selects
Stop Test. Data will be collected according to the Data Collection setting (see
Section III.5).
TIP: An End Condition of zero Time can be useful when measuring a new
material. During the test, the speed can be changed without ending and