Brookfield DV2T Viscometer User Manual
Page 59
Brookfield Engineering Labs., Inc.
Page 59
Manual No. M13-167-B0614
A.2 Setup
1. Be sure that the Viscometer is securely
mounted to the Laboratory Stand, leveled and
zeroed with no cone or cup attached and 0%
torque is displayed.
2. Figure A-2
shows a typical water bath setup.
Connect the sample cup inlet/outlet ports to the
water bath inlet and outlet and set the bath to
the desired test temperature. Allow sufficient
time for the bath to reach the test temperature.
3. The Viscometer has been supplied with a
special cone spindle(s) which contains the
Electronic Gap Setting feature. The “CPE” or
“CPA” part number designation on the cone
spindle verifies the Electronic Gap Setting
feature.
Note
:
The “CPE” or “CPA” cone or
cup cannot be used with earlier
DV2T cone/plate Viscometers
(below S/N50969) which do not
have the electronic gap setting
feature.
4. With the motor off, thread the cone spindle
by using the spindle wrench to secure the
viscometer coupling nut (see Figure A-3);
gently push up on the coupling nut and hold
this securely with the wrench. Thread the cone
spindle by hand.
Note
:
Left Hand Threads.
5. Attach the cup, taking care not to hit the cone
with the cup (Figure A-4), by positioning the
cup against the mic ring and swinging the
tension bar under the cup. The tension bar
should have plastic tubing in place.
Bath/Circulator
Bath
Inlet
Bath
Outlet
Sample
Cup
Cup
Outlet
Cup
Inlet
Figure A-3
Spindle
Wrench
Cone Spindle
These surfaces
must be clean!
Coupling Nut
Micrometer
Adjustment
Ring
Do Not hit the
CONE with the CUP!
Tension Bar
Figure A-4
Figure A-2