IV. SETTINGS
The Settings menu provides access to the many controls and features of the DV2T Viscometer.
This menu can be accessed through the Settings Navigation Icon
which is often present
in the Title Bar. Figure IV-1 shows the Settings Menu which is divided into: Device Setup,
User Settings, Global Settings, and Admin Functions. Global Settings include items that affect
the complete range of features within the DV2T. Admin Functions include items related to
administrator level controls.
Figure IV-1
IV.1 Device Setup
The Device Setup menu includes settings related to the mechanical systems and connections
of the DV2T. Six settings are available including:
Temperature - Create off sets to be used with specific temperature probes.
Printer Setup
- Define the print format to be used with the Dymo 450 Printer.
AutoZero
- Force the DV2T to perform an AutoZero as is done at start up.
Oscillation Check - Evaluate the performance of the DV2T lower bearings.
Technical Support Info - Information that may be requested by Brookfield Technical
Support staff to assist with troubleshooting.
About
- Present basic information about the DV2T as is done at start up.
Temperature:
The Temperature setting allows the user to create up to 10 temperature
offset values for connected temperature probes. Brookfield offers several
types of temperature probes for use with the DV2T including: DVP-94Y
immersion probe (supplied with the DV2T), CPA-44PYZ embedded probe
(optional with Cone/Plate DV2T Viscometers), and SC4-13RP embedded
probe (optional with Small Sample Adapter). Any of these probes can be
calibrated locally against a standard reference thermometer to determine
an offset (how far from the actual temperature does the probe read).
This offset can be entered into the DV2T Viscometer and identified with
a name defined by the user.