Brookfield DV2T Viscometer User Manual
Page 34
Brookfield Engineering Labs., Inc.
Page 34
Manual No. M13-167-B0614
Figure III-10
The More/Less Bar can be used to reduce the number of parameters shown in the display.
The Speed parameter is active in this view. The operator can change the speed of test
without returning to the Configure Viscosity Test screen.
TIP: If the speed is changed during the execution of a saved test, then the test status
will be changed to “Unsaved Test”. This will also be reflected if the collected data
is saved.
View Test includes the Stop Test command key and a Back Arrow navigation key. The
Back Arrow is used to return the display to the Running Viscosity Test view.
III.9 Results
Measurement data is viewed in the Results screen. This screen is presented at the
conclusion of a test or when data is loaded through the View Results selection from the
Home Menu.
The DV2T utilizes a comprehensive data format. Data files include the complete set of
measurement results and calculated values along with the test protocol. All elements of
the test can be viewed in the Results screen.
The DV2T Viscometer allows for 5000 total data points per file. When viewing large data
files, additional time is required when moving from the various Results options listed below.
There may be some delay on the screen while the DV2T prepares the data.