Brookfield DV2T Viscometer User Manual
Page 56
Brookfield Engineering Labs., Inc.
Page 56
Manual No. M13-167-B0614
Spindle: The spindle to be used with this
Viscosity Test. All calculations will be
based on this spindle number.
Report Screen: Select the view to be presented upon
the conclusion of the Viscosity Test.
Save Results Path: Select if a path for saving data is to be
defined. Define the save path.
Post Test Averaging: Select the type of averaging to perform
at the end of the test.
Instructions: Document the instructions to present
to the operator when the test is run.
The Test Parameters section includes all Viscosity Test program controls that may vary
if there is more than one test step.
Speed: Speed of rotation of the spindle.
Data Collection: Data Collection method.
Time Interval: Time interval used for Data Collection
methods that include averaging.
Density: Density value to be used if kinematic
viscosity units are selected (cSt, mm2/s).
End Condition: Condition which sets the end of the
Viscosity Test.
End Condition Value: Parameter based on chosen End
Condition.
QC Limits: Parameter to be used for QC limit
indicators.
QC High / QC Low Limit Parameter based on the chosen QC
Limits.
Collect Single Point at Step End: During a Multi Point data collection,
force the DV2T to collect a data point
at the end of the step (when the End
Condition is achieved).
Include This Step into Average
Calculation:
Data from this step will be included in
the average calculation indicated by the
selection in Post Test Averaging.
Add, Insert, Delete Step Buttons
Save Button
The Program Grid shows all Test Parameters for each step in a line view.
All measurement units can be set through the Units Setup button, see Figure V-2. This
menu can be pinned to the screen. If the menu is not pinned, then it will minimize when
the mouse is moved outside of the window.