A specific user can change their own password at any time. The
user must first enter the current password. Then the user will be
prompted to enter and confirm the new password.
TIP: If the administrator password is lost, it can be reset. Please
contact Brookfield or your Brookfield representative. Remember
to have the information found in the About screen available
(see Section IV.1).
IV.3 Global Settings
The Global Settings menu includes settings that affect the operation of the DV2T across
all users and menus. These settings are independent of user log in. The five Global
Settings include:
Measurement Units
- Select the unit of measure for several parameters.
Regional Settings - Specify Language, number, time, and date formats.
Global Alarm - Select a single measurement parameter for alarm status.
Spindle List
- Configure the display of available spindles; create a special
spindle.
Speed List
- Configure the display of available speeds; create a new speed.
Measurement Units:
The Measurement Units menu displays the current selections for
measurement units. Each unit can be changed by pressing the blue
down arrow and selecting the desired value. Measurement Units
are part of the data file for saved results. Available units include:
Viscosity cP
centipoise
P
Poise
mPa
•s
milliPascal seconds
Pa
•s
Pascal seconds
cSt
centistokes
mm
2
/s
millimeters squared per second
1000 cP = 10
P = 1000
mPa•s = 1 Pa•s
1 cSt = 1 mm
2
/s
TIP: When cSt units are chosen for viscosity, a
density value must be entered in the Density
field in Configure Viscosity Test.
Speed
RPM
revolutions per minute
1/s
reciprocal seconds
The relationship between RPM and 1/s is defined by
the SCR value of the selected spindle. See Appendix
D for details.