Brookfield DV2T Viscometer User Manual
Page 26
Brookfield Engineering Labs., Inc.
Page 26
Manual No. M13-167-B0614
Figure III-3
Figure III-4
Single Point
: Collect only a single data point when the End Condition is met.
Single Point Averaging
: Specify an amount of time over which to average measured
data. Collect a single data point when the End Condition is
met. This data point is an averaged value. If the time for
averaging is shorter than the total time for the step, then the
average will be performed for the specified time at the end
of the test.
Example 1: The End Condition is Time with a value of 1 minute and 30 seconds,
the Single Point Averaging Duration is 30 seconds, the single data point
collected from this step will be an average of the data measured from 1
minute to 1 minute 30 seconds.