Emc compliance – Welch Allyn Propaq Encore Monitor - User Manual User Manual
Page 20
12
English Update: Propaq Encore Monitor
Welch Allyn Propaq Encore Monitor
EMC Compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment.The Propaq Encore complies with IEC EN 60601-1-2:2001.
•
All medical electrical equipment must be installed and put into service in accordance
with the EMC information provided in this document and the Propaq Encore
Directions for Use or Propaq Encore Reference Guide.
•
Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.
Propaq Encore monitors comply with all applicable and required standards for
electromagnetic interference.
•
It does not normally affect nearby equipment and devices.
•
It is not normally affected by nearby equipment and devices.
•
It is safe to operate the monitor in the presence of high-frequency surgical
equipment.
•
However, it is good practice to avoid using the monitor in extremely close
proximity to other equipment.
Propaq Encore Monitor: Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The Propaq Encore monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the monitor should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1
The monitor uses RF energy only for its internal
function.Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The monitor is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies