This section describes normal ranges for Spot Vital Signs LXi.

Blood Pressure Accuracy

Blood pressure accuracy meets or exceeds AAMI SP10:2002 standards for non-invasive
blood pressure accuracy (± 5 mmHg mean error, 8 mmHg standard deviation). Blood
pressure accuracy is validated for pressure measurement using the upper arm only.

Cuff Pressure Range

0 to 300 mmHg

Systolic Range

60 to 250 mmHg

Diastolic Range

30 to 160 mmHg

Blood Pressure Determination Time

Typical: 15 seconds

Mean Arterial Pressure Range

40 to 190 mmHg

Pulse Rate Range
(using SpO

2

determination)

25 to 240 bpm

Pulse Rate Range
(using Blood Pressure determination)

35 to 199 bpm

Pulse Rate Accuracy (using SpO

2

determination)

Without Motion: 25 to 240 bpm ± 3 digits

1

With Motion: normal physiologic range (55 to 125 bpm) ± 5 digits

Low Perfusion: 25 to 240 bpm ± 3 digits

1

Pulse Rate Accuracy
(using Blood Pressure determination)

±5.0%

Overpressure Cutoff

315 mmHg ±15 mmHg

1

Specification applies to device performance and was validated with Biotek and Nellcor simulators.