Troubleshooting
Welch Allyn Spot Vital Signs LXi
Table 14. SureTemp Plus Temperature Error Codes
Code
Description
Corrective Action
C22
Temperature time limit exceeded.
Remove probe from patient.
E0.1
Probe heater error.
Retake reading. If problem persists, replace probe.
E0.2
Thermometer probe or device malfunction. Replace probe. If problem persists, contact Technical Service.
E0.4
Probe is over temperature.
If problem persists, contact Technical Service.
E0.5
Unable to determine temperature.
Retake reading. If problem persists, replace probe.
E0.6
Probe data error.
Retake reading. If problem persists, replace probe.
E0.7
Broken thermometer probe.
Replace probe.
E0.8
Cannot read the probe’s configuration
information.
Contact Technical Service.
E0.8
Temperature module data error.
Contact Technical Service to return the device.
E0.8
Cannot read the device’s Error Log.
This problem will correct itself. If it persists, contact
Technical Service.
E0.9
Broken thermometer probe.
Replace probe.
E4.0
Internal temperature malfunction.
Retake temperature. If problem persists, contact Technical
Service.
E4.1
Internal temperature malfunction.
Retake temperature. If problem persists, contact Technical
Service.
E4.2
Internal temperature malfunction.
If problem persists, contact Technical Service.
E4.3
Internal temperature malfunction.
If problem persists, contact Technical Service.
E4.4
Temperature malfunction.
Restart device. If problem persists, contact Technical Service.
E4.5
Temperature malfunction.
Restart device. If problem persists, contact Technical Service.
E4.6
Temperature malfunction.
Restart device. If problem persists, contact Technical Service.
E4.7
Cannot initialize thermometer.
If problem persists, contact Technical Service.
E4.8
Thermometer needs to be calibrated.
Contact Technical Service.
E4.9
Probe well missing or installed improperly. Reinstall probe well.
E5.0
Temperature heater error.
If problem persists, contact Technical Service.
E5.2
Heatsink Failsafe Failure
If problem persists, contact Technical Service.
A^!
Device outside operating temperature
range.
Change ambient temperature.
Av!
Device outside operating temperature
range.
Change ambient temperature.
b^
Internal temperature malfunction.
Contact Technical Service.
bv
Internal temperature malfunction.
Contact Technical Service.
Table 15. SpO
2
Error Codes
Code
Description
Corrective Action
C8
Faulty SpO
2
sensor.
Replace sensor.
C9
SpO
2
time limit exceeded.
Remove sensor from patient.
E7
Internal SpO
2
error.
Retake reading.