General cautions – Welch Allyn Spot Vital Signs LXi - User Manual User Manual
Page 19

Directions for Use
Introduction
9
General Cautions
A caution statement in this manual identifies a condition or practice, which if not
corrected or discontinued immediately, could lead to equipment failure, equipment
damage, or data loss.
These cautions pertain to the entire Spot Vital Signs LXi device.
WARNING Tissue damage can be caused by incorrect application or duration of use of a
Nellcor OxiMax sensor. Inspect the sensor site as directed in the sensor Directions for
Use.
WARNING Do not use the sensors during magnetic resonance imaging (MRI) scanning.
Induced current could potentially cause burns. The MS board pulse oximeter may affect
the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.
WARNING Carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
WARNING Failure to cover the Nellcor OxiMax sensor site with opaque material in high
ambient light conditions may result in inaccurate measurements.
WARNING Do not use the pulse oximeter as a replacement or substitute for ECG-based
arrhythmia analysis.
Caution If the accuracy of any measurement is in question, check the patient’s
vital sign(s) with an alternate method and then check to verify the device is
functioning properly.
Caution Place the device on a secure surface or use one of the optional
mounting accessories.
Caution Do not place fluids on or near the device.
Caution It is recommended that the device is used within stated operating
temperature ranges (see “Environmental” on page 55). The device will not meet
its performance specifications if used outside these temperatures ranges.
Caution Always unplug the AC power transformer from the outlet before
moving the mobile stand to a new location.
Caution The basket has a three-pound weight limit. Take care not to exceed
this limit.