beautypg.com

Event causes and corrective actions – Welch Allyn Spot Vital Signs LXi - User Manual User Manual

Page 57

background image

Directions for Use

Troubleshooting

47

Event Causes and Corrective Actions

Note

Differences of up to 10 mmHg between manual and automatic readings are
considered normal and occur for a number of reasons including intra-patient BP
variability, observer hearing differences, and auscultatory deflation rate.

Table 16. Inaccurate Blood Pressure Readings

Possible Cause

Corrective Action and Explanation

Incorrect cuff size

Use Welch Allyn approved cuffs only.
Measure patient’s arm circumference midway between elbow and
shoulder (see “Blood Pressure Cuff Selection” on page 33 to select correct
cuff size). Use reference markings on cuff to ensure correct cuff size.

Patient’s arm position

Ensure patient’s arm is at heart level.

Arm movement during blood pressure
cycle

Keep arm still during blood pressure cycle.

Movement may cause inaccuracies from artifact.

Blood pressure taken over clothing

Take blood pressure on a bare arm.

Arrhythmia

Check for regularity of heart rate (palpate pulse or check device).

Moderate to severe heart rate irregularities may make blood pressure
difficult to measure accurately.

Change in blood pressure between
auscultatory reading and Spot LXi reading

Check blood pressure immediately before Spot LXi reading.

Blood pressure is dynamic and changing. It is normal for blood
pressure to fluctuate 5 to 10 mmHg.

Incorrect reference

Use the correct Korotkoff sound to determine diastolic blood pressure.

Many listeners incorrectly equate diastolic blood pressure with the
disappearance of sound only (phase 5). Spot LXi was developed using
the American Heart Association recommendations, which state that
phase 5 be used unless sound continues to 0 mmHg, in which case the
change in the quality of sound (phase 4) is to be used.

Deflate cuff no faster than 3 mmHg per second.

One of the major sources of error in auscultatory blood pressure
measurement is deflating the cuff too quickly. The American Heart
Association recommends deflation no faster than 3 mmHg per second.

Only use a sphygmomanometer that is calibrated.

An uncalibrated sphygmomanometer may result in inaccurate blood
pressure measurements.

Poor auscultatory sound recognition by
observer

Use high-quality stethoscope. Have a different observer check patient’s
blood pressure.

Table 17. Cuff Inflation and Deflation with No Blood Pressure Reading Displayed (or Error Code in
Display)

Possible Cause

Corrective Action and Explanation

Leak in pneumatic system

Ensure all cuff attachments are tight.
Carefully check for leaks in the blood pressure cuff, tubing, and pressure
hose attached to Spot LXi.

Arm movement during cycle

Keep arm still during blood pressure cycle.

Movement may cause inaccuracies from artifact, long cycle times, and
error message.

Cuff tubing or pressure hose movement
artifact

Do not contact cuff tubing or pressure hose during blood pressure cycle.

Movement may cause inaccuracies from artifact.