Pulse Oximetry Operation

Welch Allyn Spot Vital Signs LXi

Spot LXi measures a patient’s SpO

2

for up to 10 minutes. After 10 minutes, a C9 error

code is displayed. This error code means that the use has exceeded the 10-minute
time limit.

2.

Check sensor sites periodically to assess circulation, sensor positioning, and skin
sensitivity.

3.

Remove the sensor from the patient. The SpO

2

reading flashes for 8 seconds. If the

sensor is not reattached to the patient in 8 seconds, the SpO

2

% and pulse rate

flashing, signalling that the measurement period has ended. The device continues to
display the last SpO

2

reading.