Page 23

background image

 

P/N 20146 Rev A

Page 23

3.

If the settings are correct, try turning the vision screener off, then back on.

4.

Check on the “Home” screen to see if your network name and IP address is identified in the

upper left corner in the black bar.

NOTE: Restarting helps to reinstate the connection to the network.

5.

If the connection does not automatically reconnect, go to Tools, Network, select Security Type,

reselect the correct type (None, WEP, WPA) then select the highlighted “Ok”. This should re-
instate the network connection.

System Messages

Specific system messages may appear while operating the vision screener. If so, the information
below will assist you in identifying what the messages indicate:

Message Text

Message

Actions

Meaning

Action

Sorry for the inconvenience.

We need to reboot the

device.

Ok or

Cancel

Screening mode has encountered

a problem and the device needs

to be restarted.

Select Ok and the device will

power cycle itself. If the problem

persists, contact Welch Allyn

Support.

Problem Encountered. The

vision screener needs to

reboot.

Reboot

The system was not able to start

and has attempted to correct the

issue

Select Reboot and try again. If

problem persists, contact Welch

Allyn Support.

The screening system is not

available.

None

The Exam system did not

initialize, or is unresponsive.

Power cycle the device and try

again. If problem persists, contact

Welch Allyn Support.

The screening system is not

initialized.

None

The Exam system did not

initialize, or is unresponsive.

Power cycle the device and try

again. If problem persists, contact

Welch Allyn Support.

The vision screener has not

finished saving the previous

screening.

None

The system has detected that it is

still saving files to disk from a

previous exam.

Power cycle the device and try

again. If problem persists, contact

Welch Allyn Support.

Un-calibrated Device

None

The system has not been able to

read the device configuration file.

Power cycle the device. If problem

persists, contact Welch Allyn

Support.


For more information on troubleshooting or error messages, contact Welch Allyn’s support team