4 calibration – Rice Lake TradeRoute HL Series - Installation and Service Manual (Legal-for-Trade) User Manual
Page 34

30
Trade
Route Installation Manual
3.4
Calibration
The CALIBR submenu (under the SCALES menu, see Figure 3-6) is used to calibrate the 920i. The zero, span, and
linear calibration point displays provide a set of softkeys used specifically for calibration procedures:
+/–
Toggles to allow entry of negative or positive values.
Last Zero
Recalls the last established zero value to allow calibration without removing tests weights or product
from scale.
Calibrate
Performs calibration for the selected point.
Temp Zero
Temporarily zeroes the displayed weight of a non-empty scale. After span calibration, the difference
between the temp zero and the previously calibrated zero value is used as an offset.
Millivolts
(or
Counts
)
Toggles between display of captured A/D counts and captured millivolts values; allows entry of
calibration values in mV or counts.
07/14/2006
01:57PM
Calibrate
WZERO
WVAL
REZERO
1291
Deadload Calibration
+/-
Millivolts
Zero Count =>
Last Zero
Temp Zero
CALIBR
SCALE1
SCALES
Figure 3-6. WZERO Calibration Display
To calibrate the indicator using the front panel:
1. Place the indicator in setup mode (display reads
Scale Configuration
) and remove all weight from the scale
platform. If your test weights require hooks or chains, place the hooks or chains on the scale for zero
calibration.
2. With the
SCALES
menu highlighted, press
and select the scale to be calibrated.
3. Press
again (
GRADS
parameter highlighted).
4. Press left to highlight the
CALIBR
submenu.
5. Press
to go to zero calibration (
WZERO
6. Ensure scale is empty and press
again to show the current WZERO value.
7. Press the
Calibrate
softkey to calibrate zero. When complete, the new A/D count for the zero calibration is
displayed.
8. Press
to save the zero calibration value and go to the next prompt (
WVAL
).
9. With
WVAL
displayed, press
to show the stored calibration weight value.