Calibration
31
5.0
Calibration
The
iQUBE
can be calibrated using the
920i
front panel or
Revolution III
. The following sections describe the
calibration procedure for each of the calibration methods.
Figure 5-1. CALIBR Submenu
5.1
920i Front Panel Calibration
The CALIBR submenu (under the SCALES menu,
see Figure 5-1) is used to calibrate the
920i
and the
iQUBE
at the same time.
Three calibration methods can be used to trim the load
cells:
•
Theoretical
•
Cal-Match
•
Manual trimming
5.1.1
Theoretical Calibration
To perform a theoretical calibration of the
iQUBE
using the
920i
front panel, press the
Theoretical
Calibration
softkey. The dead load value is captured
and the cells trimmed based on the capacity and
sensitivity values entered for the cells. Scale capacity
is set at the total of the load cell capacities.
5.1.2
Cal-Match Calibration
To perform a Cal-Match calibration of the
iQUBE
using the
920i
front panel, do the following:
1. Place the indicator in setup mode (display
reads
Scale Configuration
) and remove all
weight from the scale platform.
2. With the SCALES menu highlighted, press
the
down
key to show the CALIBR parameter.
3. From the CALIBR parameter, press
down
again to show a list of configured platforms
and select the platform to be calibrated.
4. Press
down
to go to the test weight parameter
(
WVAL
). By default, the test weight value is set
to the total load cell capacity. Use the numeric
keypad to enter the test weight, then press
enter
.
5. Press the
Cal-Match
softkey to continue
WZERO
WVAL
WSPAN
WLIN
REZERO
POINT 4
POINT 5
POINT 1
POINT 2
POINT 3
VISIBLE
GRADS
FORMAT
CALIBR
…
…
CALIB
SECTIONS
CORNERS
WZERO
display
List of
platforms
PLAT x
Sections
display/tweak
Corners
display/tweak
Deadload
adjustment
weight_value
Theoretical
Calibration
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
ALGOUT
VERS
SCALE x
If
SECTION=PAIRED
mV_value
NOTE: Perform Corner
or Section-Match