6.0 Calibration
Note
While the functionality remains the same for a standard 920i HMI weight indicator, there are certain menu
choices that need to be changed in the manual that are specific to the proper functioning of the CLS-Series
Cargo Lift Scale which include calibration.
The
920i
can be calibrated using the front panel: Use the following steps to calibrate the CLS-920i Cargo Lift
Scale.
Note
The forklift tines must be in place as test weights will be placed upon them.
Use a level to ensure the forklift tines are level from front to back (pitch), and side to side (roll) prior to
calibration.
The test weight used to calibrate the CLS Cargo Lift Scale cannot be greater than 2500 lbs.
1. With the 920i indicator in normal weighing mode, press the
More =>
softkey, then the
Supervisor Menu
softkey. Enter a password if it asks for it.
2. Using the
up/down
arrows, highlight the
Zero Inclinometer
menu item. Press the
right arrow
key.
3. Ensure that the forklift tines are level both front to back and right to left and press the
Yes
softkey.
4. Press the
Exit
softkey and return to the normal weighing mode.
5. Put the indicator into setup mode by pressing the indicator’s setup switch (located on the underside of the
indicator). The indicator’s setup menu is displayed with the
SCALES
highlighted. The following menu
structure represents the menu path that is noted in the rest of the calibration sequence.
4$"-&4
4&3*"-
'&"563&
1'03.5
4&5154
%*(*0
"-(065
7&34
4$"-&
8;&30
841"/
3&;&30
87"-
841"/
7*4*#-
8-*/
$"-*#3
…
(3"%4
41-*5
'03."5
&OUFSOVNCFS
BOEQSFTT&OUFS
4$"-&
$0/'*(
$"-*#3
4&$5*0/
…
…
1-"5
1-"5
8-*/
"UUIJTQPJOU
ZPVHFUBi$BM.BUDIwTPGULFZ1SFTT
UIF$BM.BUDITPGULFZ
UIFOTFMFDU$PSOFS$BM.BUDI
XJUIBOVQBSSPXQSFTT&OUFS
Figure 6-1. Calibration Menu for the CLS-920i
6. Press the
down
arrow twice to highlight
GRADS
, then press the
left
key once to
CALIBR
.
7. Press the down key once to
WZERO
, then the right key to
CALIBR
.
8. Press the
down
key three times to display the value of the test weight that will be used. Using the numeric
keypad on the 920i, enter the value of the test weight you will be using and press the
Enter
key. Note that if
the indicator displays the error
iQube not updated
, press the
Enter
key again. If the indicator continues to
display the error message, make sure the iQube has power applied (wired connection is good, or the battery
is in place if it is a wireless version), and that the value entered for the test weight is correct.
9. Press the
up
key once to
PLAT1
.
10. Press the
Cal-Match
softkey, then the
up
key once to highlight
Corner Cal-Match
.
11. Make sure there is no weight on either fork (zero) and press the
Enter
key. The CLS system’s zero point
will then be calibrated.