SERIAL Menu
Parameter
Choices
Description
Level 2 submenus
PORT 1
PORT 2
PORT 3
PORT 4
…
PORT x
CMD
PROGIN
KEYBD
SCALE
IND SC
IQUBE
Specifies the type of data received by the port:
CMDIN:
Remote command input
PROGIN: Programmable input
KEYBD:
Remote keyboard input
SCALE:
Legal-for-trade serial scale input
IND SC:
Industrial (non-legal-for-trade) scale input
IQUBE:
iQUBE serial scale input
HIPREC: High-precision scale input
KEYBD is available only on Port 2; SCALE, IND SC, IQUBE, and HIPREC are available only on
Ports 3 and 4 and higher (expansion ports).
NOTE: The keyboard interface is not hot-pluggable. Power-off the
920i
before plugging
the keyboard cable into the Port 2 connector
Level 3 Submenus
Port 1–Port 32
BAUD
9600
19200
28800
38400
57600
115200
300
600
1200
2400
4800
Baud rate. Selects the transmission speed for the port.
NOTE: The maximum baud rate for ports on serial expansion cards (port number greater than
4) is 19200.
BITS
8NONE
7EVEN
7ODD
8ODD
8EVEN
Selects number of data bits and parity of data transmitted or received by the port.
STOP BITS
2
1
Selects number of stop bits transmitted or received by the port.
ECHO
ON
OFF
Specifies whether characters received by the port are echoed back to the sending unit.
RESPONSE
ON
OFF
Specifies whether the port transmits replies to serial commands.
TERMIN
CR/LF
CR
Termination character. Selects termination character for data sent from the port.
EOLDLY
0
0–255
End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is
terminated to the beginning of the next formatted serial output. Value specified must be in the
range 0-255, in tenths of a second (10 = 1 second).
HANDSHK
OFF
XONOFF
HRDWAR
Specifies whether XON/XOFF flow control characters or hardware handshaking is used.
Hardware handshaking is available only on Port 2.
ADDRESS
0
0–255
Specifies the decimal indicator address for RS-485 connections. RS-232 communications is
disabled if an address other than zero is specified for this parameter. RS-485 addresses must
be in the range 01–255.
Table 3-5. Serial Menu Parameters