5 program menu, Program menu, E section 3.2.5 – Rice Lake SURVIVOR LV Series Livestock Ring Scale User Manual
Page 28

24
590 AG Indicator Operators Manual
3.2.5
Program Menu
PWRUPM
GO
NONE
OIML
REGULA
000000
number
CONSNU
DELAY
000000
number
CONSTU
NTEP
CANADA
1
number
UID
STNDBY
DSPHLD
SBYDLY
LSTMOD
HDELTA
80
10-99
DSPMOD
DSABLE
ENABLE
0
number
YES
NO
VERS
PROGRM
PFORMT
SERIAL
CALIBR
CONFIG
FORMAT
SMPTIM
4
1-10
Figure 3-8. Program Menu
PROGRM Menu
Parameter
Choices
Description
PWRUPM
GO
DELAY
Power up mode. In GO mode, the indicator goes into operation immediately after a brief power
up display test.
In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period repeated.
STNDBY
SBYDLY
LSTMOD
Specifies standby mode delay and whether indicator returns to the last operating mode or
performs a power-up sequence when standby mode ends. See sub-menu parameter
descriptions below.
STNDBY sub-menu parameters
SBYDLY
0
number
Standby mode delay. Specifies the number of minutes the indicator must be inactive before
entering standby mode. Valid values are 0 (off) or 1–120 minutes.
The indicator enters standby mode if no key presses, serial communications, or scale motion
occur for the length of time specified on this parameter. Set this parameter to 0 to disable
standby mode.
LSTMOD
YES
NO
Last mode. Specifies whether the indicator resumes operation in the last mode it was in before
entering standby mode (YES) or performs the standard power-up sequence (NO).
Table 3-4. Program Menu Parameters