3.2.9

Analog Output Menu

The ALGOUT menu is used only if the analog output option is installed. If the analog output option is installed,

configure all other indicator functions and calibrate the indicator (see Section 4.0) before configuring the analog

output. See Section 7.8 on page 50 for analog output calibration procedures.

XXXXXXX

XXXXXXX

ALGOUT

DIGIN

SETPNT

XXXXXXX

PROGRM

PFORMT

SERIAL

CALIBR

XXXXXXX

CONFIG

FORMAT

SOURCE

GROSS

0%

20%

OFFSET

ERRACT

FULLSC

HOLD

000000

number

MIN

10000

number

MAX

TWZERO

NET

TWSPAN

ZEROSC

VERS

MINNEG

OFF

ON

MAXNEG

OFF

ON

Figure 3-13. Analog Output Menu

Table 3-9. Analog Output Menu Parameters

ALG OUT Menu

Parameter

Choices

Description

Level 2 submenus

SOURCE

GROSS
NET

Specifies the source tracked by the analog output.

OFFSET

0%
20%

Zero offset. Selects whether the analog output supplies voltage (0–10 V) or current (4–20 mA)
output. Select 0% for 0–10 V output; select 20% for 4–20 mA output.

ERRACT

FULLSC
HOLD
ZEROSC

Error action. Specifies how the analog output responds to system error conditions.

Possible values are:

FULLSC

Set to full value (10 V or 20 mA)

HOLD:

Hold current value

ZEROSC:

Set to zero value (0 V or 4 mA)

MIN

000000
number

Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0–999 999.

MINNEG

OFF
ON

Turn ON or OFF.

MAX

010000
number

Specifies the maximum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0–999 999.

MAXNEG

OFF
ON

Turn ON or OFF.

TWZERO

Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press and hold

or

to adjust the output. Press

to save the new value.

TWSPAN

Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analog
output value. Press and hold

or

to adjust the output. Press

to save the new value.