Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.
OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than zero.
NONE allows tares to be acquired at any weight value.
OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no
load. NONE allows tares to be cleared at any weight value.
NTEP and OIML modes allow a new tare to be acquired even if a tare is already present. In
CANADA mode, the previous tare must be cleared before a new tare can be acquired.
NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as
long as the current weight is within the specified ZRANGE. In OIML mode, the scale must be in
gross mode before it can be zeroed; pressing the ZERO key in net mode clears the tare.
CONSNU
000000
number
Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.
The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.
CONSTU
000000
number
Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the consecutive number is reset by sending CLRCN digital input.
UID
1
Specifies the unit identification number (any numeric value up to six digits).
ACCUM
ON/OFF
RTZGRD
Turns the accumulator on and off. Stores the count, date and time of last accumulation.
Return to zero grads to re-arm the accumulator. Default = 0.
DATE
DATFMT
DATSEP
Allows selection of date format and date separator. See Level three parameter for descriptions.
TIME
TIMFMT
TIMSEP
Allows selection of time format and separator. See level three parameter for descriptions.
Level 3 submenus
ACCESS
DISABLE
ENABLE
Operator access to piece count mode. Specify DISABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.
SPLSIZ
10
20
50
100
5
PCWGT
Sample size. Specify the default size used for counting scale operations. Sample size can be
changed in counting mode during sample acquisition.
ON/OFF
ON
OFF
Turns the accumulator on and off.
RTZGRD
0
Number
This determines the number of grads away from 0 that it has to return to re-arm the
accumulator between weighments.
DATFMT
MMDDYY
DDMMYY
YYMMDD
Specifies the format used to display or print the date.
DATSEP
SLASH
DASH
SEMI
Specifies the date separator character.
TIMFMT
24HOUR
12HOUR
Specifies the format used to display or print the time.
TIMSEP
COLAN
COMMA
Specifies the time separator character.