Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than zero.
NONE allows tares to be acquired at any weight value.

OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no
load. NONE allows tares to be cleared at any weight value.

NTEP and OIML modes allow a new tare to be acquired even if a tare is already present. In
CANADA mode, the previous tare must be cleared before a new tare can be acquired.

NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as
long as the current weight is within the specified ZRANGE. In OIML mode, the scale must be in
gross mode before it can be zeroed; pressing the ZERO key in net mode clears the tare.

CONSNU

000000
number

Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.

The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.

CONSTU

000000
number

Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the consecutive number is reset by sending CLRCN digital input.

UID

1

Specifies the unit identification number (any numeric value up to six digits).

ACCUM

ON/OFF
RTZGRD

Turns the accumulator on and off. Stores the count, date and time of last accumulation.

Return to zero grads to re-arm the accumulator. Default = 0.

DATE

DATFMT
DATSEP

Allows selection of date format and date separator. See Level three parameter for descriptions.

TIME

TIMFMT
TIMSEP

Allows selection of time format and separator. See level three parameter for descriptions.

Level 3 submenus

ACCESS

DISABLE
ENABLE

Operator access to piece count mode. Specify DISABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.

SPLSIZ

10
20
50
100
5
PCWGT

Sample size. Specify the default size used for counting scale operations. Sample size can be
changed in counting mode during sample acquisition.

ON/OFF

ON
OFF

Turns the accumulator on and off.

RTZGRD

0
Number

This determines the number of grads away from 0 that it has to return to re-arm the
accumulator between weighments.

DATFMT

MMDDYY
DDMMYY
YYMMDD

Specifies the format used to display or print the date.

DATSEP

SLASH
DASH
SEMI

Specifies the date separator character.

TIMFMT

24HOUR
12HOUR

Specifies the format used to display or print the time.

TIMSEP

COLAN
COMMA

Specifies the time separator character.