41
Table A.1. Running conditions for SDS-PAGE in the Criterion cell. Do not run different gel formulations at the same time.
TGX Tris-HCl
Bis-Tris
Tris-Acetate
Running buffer
Tris/glycine/SDS
Tris/glycine/SDS
XT MOPS
XT MES
XT Tricine
Standard Conditions
Power conditions
200 V constant
200 V constant
200 V constant
200 V constant 150 V constant
Expected current (per gel)
Initial
55–80 mA
90–120 mA
165–175 mA
185–200 mA
170–180 mA
Final
33–43 mA
35–55 mA
60–70 mA
90–110 mA
85–95 mA
Run time
42–45 min
50–55 min
60 min
45 min
65 min
High Voltage (Rapid) Conditions
Power conditions
300 V constant
—
—
—
—
Expected current (per gel)
Initial
89–135 mA
—
—
—
—
Final
66–99 mA
—
—
—
—
Run time
20–26 min
—
—
—
—
Table A.2. Standard transfer (blotting) conditions.
Method
Membrane Time Voltage
Tank blotting
Nitrocellulose (0.45 μm)
20–60 min
100 V
1
PVDF (0.2 μm)
Semi-dry blotting
Nitrocellulose (0.45 μm)
20–60 min
25 V
PVDF (0.2 μm)
1
For previously optimized protocols for proteins <30 kD, reduce transfer time by 20%.
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels