2 peptide gel staining, 3 ief gel staining – Bio-Rad Criterion™ TBE-Urea Precast Gels User Manual
Page 37

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
31
Stain
Sensitivity
(Lower
Limit)
Optimum
Protein Load
(µg/Band)
Advantages
Disadvantages
Imaging
Manual
SYPRO
Ruby protein
gel stain
1–10 ng
~0.1
Broad dynamic
range
Requires laser-
or LED-based
imaging
instrument for
maximum
sensitivity
Fluorescence
visualization
with UV,
LED, or laser
scanning
4006173
Flamingo
™
fluorescent gel
stain
0.25–0.5 ng
~0.02
Broad dynamic
range, mass
spectrometry
compatible
10003321
Criterion TGX Stain-Free and Criterion Stain Free Gels
Stain Free
imaging
2–28 ng
~0.5
Rapid (<5 min),
compatible with
blotting and mass
spectrometry,
simple protocol
with no additional
reagents
Requires
tryptophan
residues in
proteins for
detection
Fluorescence
visualization
using Criterion
Stain Free
imaging system
10014472
12.2 Peptide Gel Staining
Peptides and small proteins are prone to diffusion and loss during staining. The following protocol
includes a fixing step prior to staining to prevent sample loss and is suitable for detection of bands as
low as 10–20 ng.
Fixative solution
40% methanol, 10% acetic acid
Stain solution
0.025% (w/v) Coomassie Blue G-250, 10% acetic acid
Destain solution
10% acetic acid
Place gels in fixative solution and equilibrate for 30 min. Stain gels with stain solution for 1 hr. Stain
should be used only once; reuse may result in loss of sensitivity. Destain gels three times for 15 min or
until the desired background is achieved. Some peptides may not be completely fixed and may diffuse
out of the gels if fixing and staining times are greatly exceeded.
12.3 IEF Gel Staining
Samples on IEF gels can be detected using multiple methods. Use Table 12.2 as a guide to selecting an
appropriate staining method.
Table 12 .2 . IEF gel detection methods .
Method
Sensitivity
(Lower Limit)
Optimum
Protein Load
(µg/Band)
Advantages
Disadvantages
Silver Stain Plus kit
0.6–1.2 ng
~0.01
Sensitive, robust, mass
spectrometry compatible
Requires TCA fixation
Silver stain
0.6–1.2 ng
~0.01
Stains complex proteins
(glyco- or lipoproteins)
Not mass spectrometry
compatible
SYPRO Ruby protein
gel stain
1.0–10 ng
~0.1
Broad dynamic range, mass
spectrometry compatible
Requires laser- or LED-
based imaging instrument
for maximum sensitivity
Flamingo fluorescent
gel stain
0.25–0.5 ng
~0.2
Broad dynamic range, mass
spectrometry compatible
Requires laser- or LED-
based imaging instrument
for maximum sensitivity
Instruction Manual and Application Guide
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels