4 sample preparation, 5 running conditions – Bio-Rad Criterion™ TBE-Urea Precast Gels User Manual
Page 30

24
Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
8.4 Sample Preparation
1. Determine the appropriate protein concentration of your sample based on the detection method
and load volume used. (See Chapter 12 for approximate stain sensitivities.)
2. Dilute 1 part sample with 1 part sample buffer. Do not heat the samples.
8.5 Running Conditions
Power conditions
125 V constant
Starting current
90–120 mA/gel
Final current
35–55 mA/gel
Run time
90 min
Criterion Precast Gels
This manual is related to the following products:
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels