Appendix a: quick start guide, Quick start guide – Bio-Rad Criterion™ TBE-Urea Precast Gels User Manual
Page 46

Quick Start Guide
A
Running buffer (1x)
Dilute 100 ml 10x stock (catalog #161-0732) with 900 ml diH
2
O.
Sample buffer
Use Laemmli sample buffer (catalog #161-0737).
Prepare Buffers
Prepare and Load Samples
Prepare Gels and
Assemble Electrophoresis Cell
Perform Electrophoresis
Remove the comb and tape from the gels and assemble the electrophoresis cell.
Fill the inner and outer buffer chambers with running buffer.
Fill the lower buffer chamber with 400 ml running buffer (to the fill mark).
40
Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
The following instructions are for electrophoresis of Criterion
™
precast gels using the Criterion system.
Component
Reducing
Nonreducing
Sample
5 μl
5 μl
Laemmli sample buffer
(catalog #161-0737)
4.75 μl
5 μl
β-Mercaptoethanol
0.25 μl
—
Total volume
10 μl
10 μl
Heat samples at 90–100°C for 5 min or at 70°C for 10 min.
Load the appropriate amount of sample on the gel.
Connect the electrophoresis cell to the power supply and perform electrophoresis according to
the conditions in Table A.1.
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels