6.4 Sample Preparation
1. Determine the appropriate concentration of sample to load (depends on the load volume and the
detection method used; see Chapter 12 for approximate stain sensitivities).
2. Dilute the sample with at least an equivalent volume of sample buffer (catalog #161-0739) and
reducing agent (
β-mercaptoethanol, for example). Heat the diluted sample at 90–95°C for 5 min,
or at 70°C for 10 min.
For example, combine:
5 μl sample
4.75 μl Tricine sample buffer (catalog #161-0739)
0.25
μl
β-mercaptoethanol (catalog #161-0710)
10
μl
total
volume
6.5 Running Conditions
Power conditions
125 V constant
Starting current
140–150 mA/gel
Final current
60–70 mA/gel
Run time
120 min
Criterion Precast Gels
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels