9.4 Sample Preparation
9.5 Running Conditions
Table 9.1. Running conditions for nondenaturing nucleic acid PAGE with Criterion gels in the Criterion cell.
5% and 10% Gels
15% and 4–20% Gels
Power conditions
100 V constant
150 V constant
Expected current (per gel)
Initial
20–25 mA
27–35 mA
Final
14–18 mA
20–35 mA
Run time
90 min
90 min
Criterion Precast Gels
This manual is related to the following products:
- Criterion™ TBE Precast Gels Criterion™ Cell Criterion Precast Gels Criterion Dodeca Cell 2-D Electrophoresis Workflow Criterion™ IEF Precast Gels Criterion™ Zymogram Precast Gels Criterion™ XT Tris-Acetate Precast Gels Criterion™ Tris-Tricine Precast Gels Criterion™ XT Bis-Tris Precast Gels Criterion Stain Free™ Tris-HCl Gels Criterion™ Tris-HCl Precast Gels Criterion™ TGX Stain-Free™ Precast Gels Criterion™ TGX™ Precast Gels