Shutdown, 1 daily shutdown – Bio-Rad S3™ Cell Sorter User Manual
Page 89

Instruction
Manual | 79
Shutdown
8
8.1 Daily Shutdown
The S3
™
and S3e
™
System should be shut down at the end of the day. This will turn off all
droplet formation and charging, lasers, and streams. A flush will be completed with the DI
water container to rinse the system.
IMPORTANT! Always follow the personal protective equipment (PPE) guidelines
relevant to your laboratory’s safety procedures for dealing with ethanol or bleach.
Note: At any time, the shutdown procedure can be canceled by clicking the X icon on the
dialog windows.
Note: If a user chooses to log off from the software, the force cleaning procedure option can
be run. Similar to the shutdown procedure, windows will appear with the exception of the
automatic startup window.
To shut down:
1. Ensure that the sample loading stage is in the wash position.
2. On the Setup and Maintenance tab select Shutdown. The software will confirm with a
dialog box that you intend to shut down the system (Figure 88).
Fig. 88. System shutdown check.