Troubleshooting, Troubleshooting 111 – Bio-Rad S3™ Cell Sorter User Manual
Page 121

Instruction
Manual | 111
Troubleshooting
11
Table 30 describes possible errors and causes with troubleshooting steps.
Table 30. Troubleshooting table.
Error
Possible Causes
Troubleshooting Steps
Bubble detected
Bubble in sample line
Place the loading dock in the wash position
to backflush the sample line and remove
the bubble
Nozzle door interlock
Nozzle door is open
Close nozzle door
Sort door interlock
Sort door is open
Close sort door
Waiting for QC
QC has not been run since the
Run QC to proceed with sorting
system has been started up
Not maintaining droplets
Large shift in drops
Perform Drop Delay to put the system back
into maintaining droplets
Cannot unlock loading stage
System is still washing
Wait for wash to complete
Sample is currently running
Stop sort or acquisition
QC will not pass
Calibration beads diluted
Use the undiluted ProLine
™
beads directly
from bottle
Using wrong calibration beads
Use only ProLine beads
QC criteria changed
Refer to the correct bead lot file
Fluidic error message
Quick disconnects are not
Recheck connections. An audible click must
connected properly
be heard for proper connection
Stream coming out at an angle
Clog in nozzle tip
Use Swap Tip Wizard and clean nozzle tip
Reseat nozzle tip on to the nozzle
Manual compensation
Compensation checkbox
Check and uncheck compensation
not working
not checked
box on the axis
Cannot see events labeled
Filter configuration may
Check filter configuration against
with fluorophores
be incorrect
fluorophore combination
Laser turned off
Check that both lasers are powered
Cannot see rare events
Total events displayed is
Increase cycle mode value to display
too low
more events
continues