Safety and Regulatory Compliance
This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards (Table 4).
S
AFETY
C
OMPLIANCE
This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards:
•
IEC 61010-1:2001 Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
•
IEC 61010-2-010:2003 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use, Part 2-010: Particular Requirements for
Laboratory Equipment for the Heating of Materials
•
CAN/CSA-C22.2 No. 61010-1-04 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use, Part 1: General Requirements
•
CAN/CSA-C22.2 No. 61010-2-010-04 Safety Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use, Part 2-010: Particular Requirements
for Laboratory Equipment for the Heating of Materials
•
EN 61010-1:2001 Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
•
EN 61010-2-010:2003 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use, Part 2-010: Particular Requirements for
Laboratory Equipment for the Heating of Materials
•
UL 61010-1:2004 (R2008) Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory use, Part 1: General Requirements
E
LECTROMAGNETIC
C
OMPATIBILITY
(EMC)
•
IEC61326-1:2005 Electrical Equipment for Measurement, Control, and Laboratory
Use - EMC Requirements, Class A
•
EN61326-1:2006 Electrical Equipment for Measurement, Control, and Laboratory
Use - EMC Requirements, Class A
•
FCC Part 15, Subpart B, Sections 15.107 and 15.109 as a Class A digital device
FCC W
ARNINGS
AND
N
OTES
•
Warning. Changes or modifications to this unit, not expressly approved by the party
responsible for compliance, could void the user’s authority to operate the equipment
•
Note. This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
Table 4. Conditions for safe use
Usage Aspect
Conditions for Safe Use
Rated input power
100–240 VAC, 50–60 Hz
Fuses
10 A, 250 V, 5 x 20 mm, fast blow (qty. 2)
Environment
Indoor use only
Temperature
15-31
o
C
Relative humidity
Up to 80% (noncondensing)
Altitude
Up to 2,000 meters above sea level
Overvoltage category
II
Pollution degree
2