4. Enter the incubation temperature. Press ENTER to accept the default incubation

temperature of 75°C, or use the numeric keys to type a new incubation temperature
between 0 and 100°C.

NOTE: Press PAUSE (.) to enter a decimal point. To delete a number, press
CANCEL.

In Figure 13, the incubation temperature is 95°C.

Figure 13. Enter the incubation temperature.

5. Press ENTER to start the incubation.

TIP: During incubation, use any options in the main menu except RUN. You cannot
start a run on a block that is running an incubation.

6. To stop an incubation, press CANCEL.

7. Select YES to stop the incubation or NO to continue the incubation. Press ENTER to

confirm the selection and return to the main menu.

8. Press SCREEN three times to view the incubation parameters summarized in the LAST

RUN screen.