5 peristaltic pump, 6 suppressor module «msm ii, 7 leak detector – Metrohm 861 Advanced Compact IC User Manual
Page 97: 8 rs232 interface, Peristaltic pump, Suppressor module «msm ii, Leak detector, Rs232 interface

5.1 Technical data
861 Advanced Compact IC / Instructions for Use 8.861.1033
87
Minimum pressure limit
adjustable between 0.1 ... 35.0 MPa (1...350 bar),
inactive at 0 MPa
Response time: 5 pump cycles
Pump head
Pump head volumes Main
piston: 40
µ
L
Priming piston:
20
µ
L
Pump displacement
volumes
Main piston:
28.5
µ
L
Priming piston:
14.25
µ
L
Length of stroke
Main piston:
3.6 mm
Priming piston:
1.8 mm
5.1.5 Peristaltic
pump
Type
2-channel peristaltic pump
Pump capacity
Rotational speed
20 U/min at 50 Hz
24 U/min at 60 Hz
Flow range
0.4
…
0.5 mL/min with 6.1826.110 Pump tubing
Maximum error
±
5 %
Maximum pressure
0.4 MPa (4 bar)
Pumpable liquids
Clear liquids with no solid contents
Pump tubing material
PVC
5.1.6
Suppressor module «MSM II»
Switching duration
140 ms
Pressure resistance
2.5 MPa (25 bar)
5.1.7 Leak
detector
Type
Detector with 2 electrodes approx. 1 mm above
base of interior
Response level
Resistance < 1 M
Ω
(for deion. water)
interface
Connector
Dsub 9 pin (male)
Function
TxD and RxD signal for connection with software
handshake
Default settings
9600 baud, 8 bit, 1 stop bit, no parity, XON/XOFF