2 connection of syringe and aspirating tubing, 4 connection of the drain tube for bottle rack, 5 connection of peek capillaries – Metrohm 861 Advanced Compact IC User Manual
Page 28: Connection of syringe and aspirating tubing, Connection of the drain tube for bottle rack, Connection of peek capillaries

2 Installation
861 Advanced Compact IC / Instructions for Use 8.861.1033
18
2.3.2
Connection of syringe and aspirating tubing
For manual filling of the sample loops 27 mounted on the injection
valves, the 6.2816.020 Syringe and the PTFE aspirating tubing 29 al-
ready screwed to the valve are needed. These accessories are
mounted or adjusted as follows:
1 Connect syringe
•
Remove the plastic stopper from connection socket 3 on the
front side of the 861 Advanced Compact IC (see Figure 1).
•
Push 6.2816.020 Syringe (without needle) as far as it will go
2 Lead aspirating tubing to the outside
•
Lead the PTFE aspirating tubing 29 connected to the injec-
tion valve 28 through one of the feedthroughs 4 to the out-
side.
2.3.3
Connection of the drain tube for the inner compartment
The 861 Advanced Compact IC has a connection at the front to which a
drain tube for discharged liquids in the inner compartment can be at-
tached. Proceed as follows:
1 Connect drain tube
•
Mount 6.1816.020 Silicone tubing on connection nipple on
the lower right side.
2 Lead drain tube to collecting vessel
•
Lead the other end of the drain tube to a suitable collecting
vessel and fix in place.
2.3.4
Connection of the drain tube for bottle rack
The 861 Advanced Compact IC has a connection at the rear to which a
drain tube for discharged liquids in the bottle rack can be attached.
Proceed as follows:
1 Connect drain tube
•
Mount 6.1816.020 Silicone tubing on connection nipple 9
2 Lead drain tube to collecting vessel
•
Lead the other end of the drain tube to a suitable collecting
vessel and fix in place.
2.3.5
Connection of PEEK capillaries
For the connections between high-pressure pump and detector block
6.1831.010 PEEK capillaries (i.d. = 0.25 mm, o.d. =
1
/
16
"
) are used
which are connected using either 6.2744.010 PEEK compression fit-
tings (long) or 6.2744.070 PEEK compression fittings (short).