3.3 Settings
861 Advanced Compact IC / Instructions for Use 8.861.1033
57
Suppressor
Suppressor module «MSM II»
(inactive with version 2.861.0010)
Switch the suppressor module to the next posi-
tion. The time since the last switching of the
suppressor module is displayed in the field be-
side the
button.
IC pump
High-pressure
pump
Start high-pressure pump.
Stop high-pressure pump.
Peristaltic pump
Peristaltic
pump
(inactive with version 2.861.0010)
Start peristaltic pump.
Stop peristaltic pump.
CO
2
Suppressor
853
CO
2
Suppressor
(only displayed with version 2.861.0040)
Start 853 CO
2
Suppressor.
Stop 853 CO
2
Suppressor.
System startup values
The system startup values are sent and applied
to the 861 Advanced Compact IC each time the
system is connected, a determination is started,
or the values are sent manually with
unit>
.
Flow, mL/min
Startup value for flow rate of the high-pressure
pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
35.0 MPa
If the
Flow
is higher than 1.5 mL/min, the defined
Pressure max
must not exceed 25.0 MPa (250 bar).
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
35.0
MPa
Full scale,
µ
S/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000, 5000
µ
S/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1