3 Operation
861 Advanced Compact IC / Instructions for Use 8.861.1033
56
3.3 Settings
3.3.1
Instrument control for connected system
A double-click on the instrument icon or selection of the
Open
menu
item using the right mouse button opens the system settings window.
For a connected system, the
Control
tab is displayed. It allows manual
control of the 861 Advanced Compact IC functions and setting of start-
up values to be sent to the instrument. This tab shows the current mea-
surement values for conductivity and pressure.
Conductivity,
µ
S/cm
Live display of measured conductivity.
Pressure, MPa
Live display of measured pressure.
The color settings for this two fields can be
changed by clicking the fields with the right
mouse button and selecting the menu item
Choose color / ...
.
Actual values
Display of actual values
Flow, mL/min
Display of flow rate of the high-pressure pump.
Pressure max, MPa
Display of maximum pressure limit for high-
pressure pump.
Pressure min, MPa
Display of minimum pressure limit for high-
pressure pump.
Full scale,
µ
S/cm
Display of selected full scale range.
Remote lines
Display of current remote line settings.
Valve
Injection
valve
Switch injection valve to "INJECT" position.
Switch injection valve to "FILL" position.