3.3 Settings

861 Advanced Compact IC / Instructions for Use 8.861.1033

59

List of program instructions
The following program instructions can be added to the time program

on the

Program

subpage:

Valve Inject,

Fill

Switch injection valve to "INJECT"
or "FILL" position.

FullScale

50, 250, 1000,
5000

µ

S/cm

Set full scale range to the selected
value.

ICPump on,

off

Switch on or off the high-pressure
pump
.

Flow

0.2 ... 2.5 mL/min

Set flow rate of the high-pressure
pump to the desired value.

Pmax

0.0 ... 35.0 MPa

Set maximum pressure limit for the

high-pressure pump to the desired
value.

Pmin

0.0 ... 35.0 MPa

Set minimum pressure limit for the
high-pressure pump to the desired
value.

Remote

0, 1, *, p

Set remote output lines 1...8 to the
desired values. For entry of the first

value, enter

1

,

0

,

p

or

*

. For entry of

the other values, move the cursor in
front of the value to be changed and
enter

1

,

0

,

p

or

*

.

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the chro-
matogram duration. Additional steps
after this flag are not allowed. The

RESET

flag is used to reset the pa-

rameters to the system startup val-
ues.

Suppressor

Switch

suppressor module «MSM

II» to the next position.

Peristaltic on,

off

Switch on or off the peristaltic
pump
.