3.3 Settings
861 Advanced Compact IC / Instructions for Use 8.861.1033
59
List of program instructions
The following program instructions can be added to the time program
on the
Program
subpage:
Valve Inject,
Fill
Switch injection valve to "INJECT"
or "FILL" position.
FullScale
50, 250, 1000,
5000
µ
S/cm
Set full scale range to the selected
value.
ICPump on,
off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5 mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 35.0 MPa
Set maximum pressure limit for the
high-pressure pump to the desired
value.
Pmin
0.0 ... 35.0 MPa
Set minimum pressure limit for the
high-pressure pump to the desired
value.
Remote
0, 1, *, p
Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter
1
,
0
,
p
or
*
. For entry of
the other values, move the cursor in
front of the value to be changed and
enter
1
,
0
,
p
or
*
.
Program
END, RESET
The
END
flag can be used to end a
program, especially if the program
time should be longer than the chro-
matogram duration. Additional steps
after this flag are not allowed. The
RESET
flag is used to reset the pa-
rameters to the system startup val-
ues.
Suppressor
Switch
suppressor module «MSM
Peristaltic on,
off
Switch on or off the peristaltic
pump.